Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Astro-Med MT95K2 V Ref (1 year) CAL VER /34401 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 47 CONFIGURATION: HP 34401A ============================================================================= # # Source: # Astro-Med MT95K2 Verification Checklist # # Compatibility: # 5500/CAL 4.11 or MET/CAL 4.11 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 34401A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD PRECISION INTERNAL VOLTAGE REFERENCE: [MEM2] AWP-1 Board 1.002 RSLT =[MEM2] AWP-1 Board 1.003 DISP Connect the 34401A to K1 on the UUT [MEM2] AWP-1 board 1.003 DISP as follows: 1.003 DISP [32] 34401A Input HI to pin 8 1.003 DISP [32] 34401A Input LO to pin 1 (signal ground) 1.004 34401 2.5000V 0.0025U 2W 2.001 HEAD PRECISION INTERNAL VOLTAGE REFERENCE: [MEM2] AWP-2 Board 2.002 RSLT =[MEM2] AWP-2 Board 2.003 DISP Connect the 34401A to K1 on the UUT [MEM2] AWP-2 board 2.003 DISP as follows: 2.003 DISP [32] 34401A Input HI to pin 8 2.003 DISP [32] 34401A Input LO to pin 1 (signal ground) 2.004 34401 5.0000V 0.0025U 2W 3.001 END